Medical Device Classification in India: Class A, B, C and D Explained

Independent private consultancy. Aaditech is not BIS or any government department and is not affiliated with or endorsed by them. Licences are granted only by the authority, and you can apply directly on its portal. Our consultancy fees are separate from official fees.

How CDSCO classifies medical devices into Class A, B, C and D under MDR 2017, which authority licenses each class, and how to find your device's class.

India sorts medical devices into four risk classes, A to D, under the Medical Devices Rules, 2017 (MDR 2017). The class decides which licence a device needs, who grants it, and how much evidence the application has to carry. Getting it right is the first step of any CDSCO medical device registration.

The four classes

ClassRisk levelIndicative examples
ALowTongue depressors, absorbent cotton, non-sterile examination products
BLow to moderateHypodermic needles, suction equipment, clinical thermometers
CModerate to highVentilators, infusion pumps, bone fixation plates and screws
DHighHeart valves, coronary stents, implantable defibrillators

Examples are indicative only. The same kind of product can sit in different classes depending on its intended use, so always check the specific device.

In-vitro diagnostic (IVD) devices are also classified A to D, using separate rules based on the risk to the patient and to public health.

How the class is decided

The rules are set out in the First Schedule of MDR 2017 and follow the approach used internationally by the GHTF and IMDRF. The main factors are:

  • Intended use stated by the manufacturer, which matters more than what the device looks like
  • Duration of contact: transient, short term or long term
  • Invasiveness: whether it enters the body through an orifice, surgically, or is implanted
  • Active or not: whether it depends on electrical or other energy
  • Body part: devices in contact with the heart, central circulatory system or central nervous system are classed higher
  • Medicinal substances or animal tissue in the device, which push the class up

Where more than one rule applies, the higher class is used.

CDSCO’s classification lists

CDSCO has published risk classification lists for many device categories, such as cardiology, orthopaedics, dental, ophthalmic and general hospital devices. These are on the medical devices section of cdsco.gov.in. If a device appears in a list, that class applies. If it doesn’t, the class is worked out from the First Schedule rules, and the reasoning should be set out in the Device Master File.

Why the class matters

Class AClass BClass CClass D
Manufacturing licenceMD-5 from the State Licensing AuthorityMD-5 from the State Licensing AuthorityMD-9 from CDSCOMD-9 from CDSCO
Import licenceMD-15 from CDSCOMD-15 from CDSCOMD-15 from CDSCOMD-15 from CDSCO
Site checkNotified body audit within 120 days after grantNotified body audit before grantCDSCO inspection before grantCDSCO inspection before grant
Licensing for devices brought under the rules in 20201 October 20221 October 20221 October 20231 October 2023

One group of Class A devices is treated differently: those that are non-sterile and non-measuring do not need a manufacturing or import licence. Since October 2022 they are registered on CDSCO’s online portal instead, with no government fee, although labelling rules still apply.

A higher class also means a heavier dossier: more test data, and for new devices without a predicate, possibly clinical investigation. Importers should read our page on the medical device import licence (MD-14/MD-15).

Do medical devices need BIS too?

Generally, CDSCO is the regulator for medical devices, not BIS. A few medical products are also covered by BIS Quality Control Orders, including diagnostic medical X-ray equipment and medical textiles. X-ray equipment is also regulated by the Atomic Energy Regulatory Board (AERB) through its eLORA portal. Wireless devices may need WPC approval, and electrical and electronic equipment can fall under the e-waste EPR rules. Check each one for your product.

Not sure of your device’s class? Aaditech offers a free first consultation to confirm the class and the licence route before you prepare a dossier.

Frequently asked questions

How many classes of medical devices are there in India?

Four. Under the Medical Devices Rules, 2017, devices are Class A (low risk), Class B (low to moderate risk), Class C (moderate to high risk) or Class D (high risk). In-vitro diagnostic devices use the same four classes with their own rules.

Where can I find the class of my device?

CDSCO publishes risk classification lists by device category on cdsco.gov.in. If your device is not listed, its class is worked out from the classification rules in the First Schedule of MDR 2017.

Does the class decide who grants the licence?

For manufacturing, yes: the State Licensing Authority licenses Class A and B devices and CDSCO licenses Class C and D. Import licences for every class are granted by CDSCO.

What if a device could fall into two classes?

The higher class applies. The same is true when a device is used with an accessory, or when more than one classification rule fits.

This article is general guidance based on official sources at the time of writing. Regulations, fees and timelines change. Check the latest notification or speak to our team before acting.